7 upgrade of the instrument software, Standard operations – Hach-Lange DR2800 eco User Manual

Page 36

Advertising
background image

36

Standard Operations

4. After "Barcode controlnumber?", a data overview is displayed,

including the specifications of the test that is to be revised.

Note: The working procedures of the corresponding test contains the
new data (Wavelength, Factors, Measurement Ranges, Conversion
Factor for chemical Form, etc.).

5. Highlight the corresponding line and press Edit.

6. When finished programming, press Store and then Cancel to

return to the User Programs menu.

7. Press Start to begin the test with the new data.

Option 2: Manual check/revision of test data
If an incorrect entry was made during the data update and is not
recognized via the barcode of the cuvette (e.g. an incorrect
measuring range), another option is available for checking and if
necessary revising, the test data.

1. Insert the light shield in compartment #2.

2. The instrument need to be at the Main Menu or press Barcode

Programs in the Main Menu.

3. Insert the blank or sample cuvette (depending on the working

procedure) into the cell compartment.

4. Press Options>More>Edit.

5. A data overview is displayed, including the specifications of the

test that is to be revised. Compare the displayed data with the
data in the working procedure and edit the displayed data in
accordance with the working procedure.

5.4.7 Upgrade of the instrument software

To obtain the software for the update from the Internet at
www.hach-lange.com:

1. Go to http://www.hach-lange.com.

2. Select the country and go to Download>Software.

3. Enter DR2800 eco in ”Search for documents“.

4. Locate the appropriate download and follow the prompts for

saving the file(s) to the USB memory stick (

Section 9 on page

57

) or to the PC.

5. Unpack the ZIP file and save the files to the USB memory stick.

6. Press Instrument Update in the System Checks menu.