Advanced Operations

10. Insert sample cuvette/cell into the cell compartment and press

Read.

11. 5 Replicates of blanking and reading lead to results.

12. The result is displayed and is to be compared with the

nominal/standard data of the sample cuvette/cell.

13. Press Cancel to return to Optical Checks.

6.1.3.4 Verification kit

The Verification Kit (see

Section 9 on page 123

) is designed for

periodic monitoring of scattered light, photometric accuracy and the
wavelength accuracy of the spectrophotometers.

When results exceed allowable tolerances (given in the quality
control certificate to the test record), contact the manufacturer.

1. Press Verification Kit in the Optical Checks menu.

2. Press Nominal Values.