4 output checks, 5 lamp history, 4 output checks 6.1.5 lamp history – Hach-Lange DR2800 eco User Manual

Page 45

Advertising
background image

45

Advanced Operations

8. Insert the different filter in the given order one after the other.

Press Next after inserting a filter.

After the last measurement the results are displayed.

9. Press PC & Printer icon to send the data to a USB memory

stick, PC or to a printer.

The files will be stored automatically as CSV file (Comma
Separated Value). The file name will be formatted as
“Verification.csv“.

6.1.4 Output checks

If a printer is connected a test printing of the current screen will be
printed.

6.1.5 Lamp history

The Lamp History menu provides the amount of time that the lamp
has been on (Hours).

After a lamp is replaced and the Lamp History is reset, the display
of the total operating time is reset to 0.

1. Press Lamp History in the System Checks menu.

2. Press Reset VIS and the Visible Lamp will be reset.

3. Press OK to return to System Checks.

Advertising