Operation

Page 36

Manual Bühler 2000

FREE PROGRAMMABLE INPUTS

Ø PROGR. INPUT
Ø ADDITIONAL INPUT 1
Ø ADDITIONAL INPUT 2
Ø ADDITIONAL INPUT 3


In the basic version 1 programmable input is available.
As an option with the "I / O expansion board" are another 3 Inputs available. Thus the
sampler can be controlled with a digital pulse (eg with external PLC).

Each INPUT (1-3) can be programmed out of the following list:

Ø NO FUNCTION

(turn off of the Input)

Ø PRG. START PULSE

(if selected, Program x will be started)

Ø PRG. STOP PULSE

(if selected, Program x will be stopped) (Back with ESC)

Ø PRG. RUN DURING PULSE

(During continuous signal, Program is running. If Signal drops down,

program stops.)

Ø DISTRIBUTOR PULSE

Pulse signal: <= 3sec means "next bottle"

>= 5 sec means "move to bottle 1"

Ø SAMPLE PULSE

Pulse means "take a sample"

Pulse signal must be > 50ms!

Remark:

This feature is only available if no program is running. The unit is

in this case completely controlled externally (eg PLC)