Hach-Lange BUHLER 1027 2000 User Manual

Page 51

Advertising
background image

Installation

Page 51

Manual BÜHLER 1027

Characteristic features: BÜHLER 1027 sampler


The device consists of the following component parts:

1. Housing/protective top

2. Suction hose connector

3. Metering vessel

4. Pinch Valve

5. Discharge tube

6. Control unit with back-lit

LC-display


7. Keypad

8. MiniUSB interface

9. Cable feedthrough for mains

cable


10. Cable feedthrough for air

inlet/outlet


11. Signal connector

Installing the apparatus


-

We recommend to install the apparatus as close to the sampling point as possible.

-

The hose has to be laid with a constant fall from the sampler to the sampling point as lower

lying points can lead to deposits in the hose which may freeze in winter or result in cross-

contaminations.

-

To fix the suction hose, we recommend to use the extraction unit available as accessory.

-

Immerse the end of the hose at the sampling point with the open end facing downstream

(direction of flow) so that coarse matter and fibres cannot be forced into the suction aperture


Mains connection

The device is equipped with a power supply lead.

The connecting data can be found in chapter “Specifications”.


Switch on / off

The device is switched on and off by means of the shock-proof plug.


Hose connection

The opening for connection of the suction hose is top right beside the dosing

unit.


Signal connections

The connector for signals (e.g. flow meter) is located at the bottom of the

control housing

Messages

Collective malfunction message (is directly connected to the relay).

Remark

:

The apparatus must be cleaned regularly in accordance with the degree of contamination present. In
view of the quality of samples, we recommend to clean thoroughly especially the wetted parts like dosing
unit, electrodes, distributor, bottles and inlet hose. Failure to do so could result in damage or destruction
to the equipment, device that

are not covered by warranty.