Specifications – Hach-Lange BUHLER 1027 2000 User Manual

Page 8

Advertising
background image

Specifications

Page 8

Manual BÜHLER 2000

Device designation

Bühler 2000

Housing

PE/PC(GF10)

Thermostatic control

Insulated sample compartment (40 mm insulating layer)
Option: freezer packs (200x10x8 mm)
Option: compressor cooling (12V/115V/230V)

Control

Microprocessor control, foil keyboard, graphical display (128*64
Pixel), back lit

Data memory

Non-volatile data memory: storage of sampling and malfunction data
like sample extractions, bottle changes, messages, external signals

Programming

12 user programs (can be edited freely)

Program start options

Immediately, at a certain time, by an external signal

Program stop options

End of sampling program after one program run, continuous operation
or x-runs

Pause mode

Interruption of program at any time

Overfilling protection

Adjustable from 1–999 samples/bottle

Interval setting

1 min. to 99 h 59 min. in steps of 1 minute

Pulse setting

1 to 9999 pulses/sample

Manual sample extraction

Possible at any time without interrupting the current program run

Program protection

Up to 5 years after voltage loss

Interface

Mini-USB, RS422/485, RS 232

Communication

Optional: TCP/IP RJ45 WEB board, 4-32GB SD/SDHC memorycard
Upon request

Languages

Multi-language, selectable

Signal inputs

• 2x analogue: 0/4-20 mA,
• 8x digital (flow, event, 1x free programmable)
Optional: expansion 4x digital, 3 free programmable and
8x Analogue Inputs 0-20mA / 0-10 V
Pulse-length min. 60 ms u. switch-level 7-24V, max. working
resistance 500 Ohm signal-length max. 30 m

Signal outputs / status
messages

Max. 8 digital outputs, 1x collective message
Optional: expansion 8x digital, 5 are free programmable

Sampling method

Vacuum system 20-350 ml

Single sample volume
accuracy

< 2,8 %

Suction height

Max. 6,5 m (at 1013hPa)

Pumping speed

>0,5 m/s at a suction height of up to at least 5 m (at 1013hPa); pump
capacity can be adjusted electronically

Suction hose

PVC, L=5 m, ID=9 mm (ID 16 mm at flow-proportional option).
Max. hose length 30 m.

Sampling modes

Time-related, flow-dependent, flow-proportional (option), event-
related, manual sample

Bottle variants

24 x 1 L PE (standard)

13 L composite container
25 L composite container
4 x 5 L PE
16 x 1 L with freezer packs

Overall dimensions

(Hxwxd)
787 x 510 x 468 mm

1030 x 548 x 470 mm with compressor cooling

Weight

22,4 kg 24x1 L version
34 kg 24x1 L version with compressor cooling (sampler incl. battery,
without suction hose, empty bottles)

Power supply

12 V/ 10 Ah lead gel battery (maintenance-free, sealed, leak proof)
115V or 230V mains-operation by means of battery charger in buffer

Advertising