Specifications – Hach-Lange BUHLER 1027 2000 User Manual

Page 50

Advertising
background image

Specifications

Page 50

Manual BÜHLER 1027

Device designation

Bühler 1027

Housing

Sampler for wall mounting
PS/PC(GF10)

Thermostatic control

Optionally available sample cabinet with self-contained cooling for
indoor installation.
Temperature in sample compartment: 4°C

Control

Microprocessor control, foil keyboard, graphical display (128*64
Pixel), back lit

Data memory

Non-volatile data memory: storage of sampling and malfunction data
like sample extractions, bottle changes, messages, external signals

Programming

12 user programs (can be edited freely)

Program start options

Immediately, at a certain time, by an external signal

Program stop options

End of sampling program after one program run, continuous operation
or x-runs

Pause mode

Interruption of program at any time

Overfilling protection

Adjustable from 1–999 samples/bottle

Interval setting

1 min. to 99 h 59 min. in steps of 1 minute

Pulse setting

1 to 9999 pulses/sample

Manual sample extraction

Possible at any time without interrupting the current program run

Program protection

Up to 5 years after voltage loss

Interface

Mini-USB, RS422/485, RS 232

Communication

Optional: TCP/IP RJ45 WEB board, 4-32GB SD/SDHC memorycard
Upon request

Languages

Multi-language, selectable

Signal inputs

• 2x analogue: 0/4-20 mA,
• 5x digital (flow, event, 1x free programmable)
Optional: expansion 4x digital, 3 free programmable and
8x Analogue Inputs 0-20mA / 0-10 V
Pulse-length min. 60 ms u. switch-level 7-24V, max. working
resistance 500 Ohm signal-length max. 30 m

Signal outputs / status
messages

Max. 8 digital outputs, 1x collective message
Optional: expansion 8x digital, 5 are free programmable

Sampling method

Vacuum system 20-350 ml

Single sample volume
accuracy

< 2,8 % at standard vacuum system

Suction height

Max. 8 m (at 1013hPa and stagnant medium)

Pumping speed

>0,5 m/s at a suction height of up to at least 4,5 m (at 1013hPa);
pump capacity can be adjusted electronically, higher with optional
diaphragm pumps

Suction hose

PVC, L=5 m, ID=12 mm. Max. hose length 30 m.

Sampling modes

Time-related, flow-dependent, flow-proportional (option), event-
related, manual sample

Bottle variants

Plastic composite container (10 L, 25 L etc.)
In combination with cooled sample cabinet:
1 x 10 L, 1 x 25 L

Overall dimensions

(Hxwxd) 362 x 442 x 222 mm
Sampler without sample container

Weight

10 kg (sampler without sample container)

Power supply

230 V / 115 V AC

Power requirement

Approx. 25 VA sampler, approx. 250 VA cooled sample cabinet

Ambient temperature

0 to + 43°C

Sample temperature

0 – 40°C

Standards

Device meets ISO 5667 standard

Wetted materials

PVC, Silicone, PS, PE, EPDM (optional: Duran 50 glass metering
vessel, stainless steel SS304 sinker weight



Advertising