Prepare the sampler, Clean the sample bottles, Install a single bottle – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS User Manual User Manual

Page 12: Clean the sample bottles install a single bottle

Advertising
background image

• The refrigerated sampler must be installed in an indoor location. The AWRS can be installed in an

outdoor location.

• Make sure that the temperature at the site location is within the specifications.
• Install the sampler on a level surface. Adjust the sampler feet to make the sampler level.
• Use the installed anchor brackets and user-supplied

3

/

8

-in. bolts for the AWRS. Refer to

Figure 3

.

• Plumb a drain tube to the ½ in.-14 NPT female connector on the bottom of the sampler.

Figure 3 AWRS anchor bracket locations with mounting dimensions

1 Optional anchor brackets

2 Anchor brackets (2x)

Prepare the sampler

Clean the sample bottles
Clean the sample bottles and caps with a brush, water and a mild detergent. Flush the containers
with fresh water followed by a distilled water rinse.

Install a single bottle
Use a single bottle to collect one composite sample. When the bottle is full, the full bottle shut-off
stops the sample program.

1. Clean the sample bottles. Refer to

Clean the sample bottles

on page 10.

2. Install the sample bottle as shown in

Figure 4

.

10 English