Plumb the sampler – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS User Manual User Manual
Page 14
Figure 5 Multiple bottle installation
1 Retainer (1322)
4 8–24 bottle tray (1511)
7 Front of sampler
2 Retainer (1056)
5 Slot for bottle tray
8 Removable tray (AWRS)
(5697600)
3 Bottle one indicator
6 Floor of refrigerated sampler
9 Insert (refrigerated sampler)
(2038)
Plumb the sampler
Install the intake tube in the middle of the sample stream (not near the surface or bottom) to make
sure that a representative sample is collected.
1. Connect the tubing to the sampler as shown in
.
Note: Use connection kit 2186 if Teflon-lined tubing is used.
2. Install the intake tube and strainer in the main stream of the sample source where the water is
turbulent and well-mixed. Refer to
.
• Make the intake tube as short as possible.
• Keep the intake tube at a maximum vertical slope so that the tube drains completely between
samples.
Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid sensor. Calibrate the
sample volume manually.
• Make sure that the intake tube is not pinched.
12 English