Figure 5 Multiple bottle installation

1 Retainer (1322)

4 8–24 bottle tray (1511)

7 Front of sampler

2 Retainer (1056)

5 Slot for bottle tray

8 Removable tray (AWRS)

(5697600)

3 Bottle one indicator

6 Floor of refrigerated sampler

9 Insert (refrigerated sampler)

(2038)

Plumb the sampler

Install the intake tube in the middle of the sample stream (not near the surface or bottom) to make
sure that a representative sample is collected.

1. Connect the tubing to the sampler as shown in

Figure 6

.

Note: Use connection kit 2186 if Teflon-lined tubing is used.

2. Install the intake tube and strainer in the main stream of the sample source where the water is

turbulent and well-mixed. Refer to

Figure 7

.

• Make the intake tube as short as possible.
• Keep the intake tube at a maximum vertical slope so that the tube drains completely between

samples.

Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid sensor. Calibrate the
sample volume manually.

• Make sure that the intake tube is not pinched.

12 English