Time based collection

1. Select or enter the values for each option.

Option

Description

Pacing interval

Collect samples at regular time intervals, in hours and minutes (0:01–999:00).

Take first sample The program starts immediately or after the first interval.

Variable volume

Specify whether the sample volumes vary (flow based). If yes, refer to

Use variable

volume

on page 21. If no, refer to

Do not use variable volume

on page 21.

Use variable volume
Sample pacing is constant time variable volume (CTVV).

1. Select or enter the values for each option.

Option

Description

Select unit

Select the flow units to use.

Map 4 mA input

Enter the flow rate that agrees with 4 mA input.

Map 20 mA input

Enter the flow rate that agrees with 20 mA input.

Average flow rate

Specify the average flow rate for the site (1–999,999). The units are specified by
the units selected above.

Time volume desired

Specify the total volume to be collected during the sample program (10–
10,000 ml).

Collection period

Specify the total collection period in hours and minutes (0:01–999:00).

Minimum sample volume If the flow volume is not sufficient to collect this minimum volume, the sample is

skipped (10–10,000 ml).

Do not use variable volume

1. Select or enter the values for the option.

Option

Description

Take first sample Select whether the program starts immediately or after the first time interval has passed.

Flow based collection

Specify the flow source for flow based sampling:

• Counts—Refer to

Use counts for the flow source

on page 21.

4–20 mA input—Refer to

Use 4–20 mA input for the flow source

on page 22.

Use counts for the flow source

1. Select or enter the values for each option.

Option

Description

Take sample every Collect samples at regular flow intervals, in counts (1–9999 counts).

Overrride time

Collect a sample if the flow volume is very low.