Configure the sdi-12 device – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS User Manual User Manual

Page 20

Advertising
background image

Option

Description

Setup base

Select the sampler base (All weather refrigerated, refrigerated or portable).

Language

Select the language of the controller.

Liq sensor cal

Calibrate the liquid sensor.

Pump duty cycle

Change the pump duty cycle. Range: 75% to 100% (default: 100%).

Thermal setup (AWRS
only)

Change the setpoint temperature inside the refrigerated compartment. Range:
2 to 10 °C (default: 4 °C).

Set contrast

Adjust the contrast of the LCD screen.

Password setup

Set up a password to control access to the program setup and the system setup
menus. Select ENABLE>YES to set a new password or ENABLE>NO to
activate the current password. If the password is set for the first time, use
900900 as the current password.

Tubing life

Set a reminder when the pump tubing is to be changed. Go to ENABLE>RESET
CYCLES to reset the pump cycle count to 0. Go to ENABLE>CYCLE LIMIT to
change the number of pump cycles that is used for the reminder. Go to
ENABLE>STATUS to show the current number of pump cycle counts.

SDI–12 (optional)

Configure the operation of an SDI–12 device. This option is shown only when a
device is found.

Configure the SDI-12 device

A sensor such as a sonde can connect to the optional SDI-12 interface. The controller can get up to
nine measurements from one SDI-12 sensor. The sensor data can then be used as a trigger in
setpoint sampling applications. Configure the parameter type and unit, and set how often the
SDI-12 device is scanned for new measurement data.

1. Select Menu>System Setup>SDI-12.

Note: The SDI-12 option is only shown when an SDI-12 device is found by the controller.

2. Select one of the options:

Option

Description

DISPLAY SONDE

Shows information about the device such as the vendor name, model name,
SDI-12 address, protocol version, number of available measurements and the warm-
up time.

CFG PARAMETERS Gives a name and measurement unit to the SDI-12 channel. Each channel is initially

shown with a generic reference indicator in the same order that is set in the
SDI-12 device.

Table 2

shows the measurements that are supported by the controller.

Note: It may be necessary to configure the SDI-12 interface for the sonde.

SCAN INTERVAL

Sets the time interval when measurement data is refreshed in the controller. Short
time intervals decrease the battery life.

Table 2 Supported measurements

Measurement

Unit

Resolution

1

Ammonia (NH3)

mg/L-N

x.xx

Ammonium (NH4+)

mg/L-N, mV

x.xxxx

Chloride (Cl-)

mg/L-N, mV

x.xxxx

Chlorophyll

µg/L, V

x.xx

Conductivity

mS/cm, µS/cm

x.xxx

Dissolved oxygen (DO)

mg/L, % saturation

x.xx

18 English