Clean the rotor

Clean the rotor, pump tube tracks and pump housing with a mild detergent. Refer to

Replace the

pump tubing

on page 36 and the illustrated steps that follow.

Replace the distributor arm tube

The distributor arm moves over each bottle during multiple bottle sampling. Replace the tube in the
distributor arm when the tube is worn. Make sure that the correct tube is used for the correct
distributor and distributor arm.

1. Remove the tube from the distributor arm and from the ceiling of the sampler cabinet.
2. Insert the new tube into the distributor arm. Make sure that the tube does not extend out of the

nozzle more than 0.3 cm (

1

/

8

in.). Refer to

Figure 17

.

3. Insert the other end of the tube into the fitting on the ceiling of the sampler cabinet.
4. Complete the diagnostic test for the distributor to make sure that the operation is correct.

Figure 17 Distributor assembly

1 Nozzle

3 Distributor motor

5 Distributor tube

2 Distributor arm

4 Shaft