Section 7

Troubleshooting

7.1 Error Codes

When a sensor is experiencing an error condition, the sensor reading on the
measurement screen will flash and all relays and analog outputs associated with the
sensor will be held.

Highlight the Probe Diag menu and press

ENTER

. Highlight Errors and press

ENTER

to

determine the case of the error. Errors are defined in

Table 3

.

7.2 Warnings

A sensor warning will leave all menus, relays, and outputs functioning normally, but will
cause a warning icon to flash on the right side of the display. Highlight the Probe Diag
menu and press

ENTER

to determine the cause of the error. Warnings are defined in

Table 4

.

A warning may be used to trigger a relay and users with the sc100 controller can set
warning levels to define the severity of the warning (see the sc100 controller user manual
for more information about warning levels). Errors are defined in

Table 4

.

Table 3 Error Codes

Displayed Error Definition

A/D Fail

Failed A/D converter. Call the Service Department.

Lamp Fail

The light source has failed. See

section 6.4.3 on page 27

for instructions for

replacing it.

Flash Fail

Datalog and event log will not work.

Table 4 Warning Codes

Warning

Number

Displayed Warning

Definition/Resolution

1

Dark Reading Warning

Dark reading detects too much light. (> 0.05 NTU)

2

Temp Warning

Sensor head internal temperature is higher than specified. Contact the Service
Department. (> 70 °C)

3

Data Log Full

Sensor data log is full. No additional data will be logged until sensor log is
downloaded into controller memory.

4

Event Log Full

Sensor data log is full. No additional data will be logged until sensor log is
downloaded into controller memory.

5

5 Volt Warn

Monitored voltage is outside the range of 4.5–5.5 V.

6

Inp V Warn

Monitored voltage is outside the range of 9.08–14.13 V.

7

Lamp Volt Warn

Monitored voltage is outside the range of 3.96–4.48 V.

8

Lamp Curr Warn

Monitored current is outside the range of 1.67–2.75 Amps.

9

Output Mode Warn

Activated when the sensor is not in normal measurement mode (such as when
in calibration or verification mode).