Specifications – Hach-Lange 1720E Low Range Turbidimeter User Manual

Page 8

Advertising
background image

4

Specifications

Verification (Wet) Method

1.

StablCal

®

(stabilized formazin) – recommended for verification in the appropriate

application range of measurement. For regulatory verification, standards of 0.1 to 50
NTU.

2.

Formazin – fresh user-prepared standard

Verification (Dry) Method

1.

ICE-PIC™ Verification Module with factory-set values of 20.0 or 1.0 ±25%. Unique value
is assigned when dry verification is done immediately after calibration and is used as
pass/fail criteria for subsequent verifications.

Recommended
Maintenance

1.

Lamp replacement: once per year.

2.

Cleaning: mandatory before calibration, optional before calibration, and mandatory upon
verification failure.

Installation Environment

Indoor

Primary Compliance
Method

USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B

Limit of Detection

1

0.0032 NTU (according to criteria specified by ISO 15839)

1

All specifications are based on a calibration with 20.0 NTU formazin and with the offset turned off.

Table 1 1720E Low Range Specifications (continued)