Specifications – Hach-Lange 1720E Low Range Turbidimeter User Manual
Page 8

4
Specifications
Verification (Wet) Method
1.
StablCal
®
(stabilized formazin) – recommended for verification in the appropriate
application range of measurement. For regulatory verification, standards of 0.1 to 50
NTU.
2.
Formazin – fresh user-prepared standard
Verification (Dry) Method
1.
ICE-PIC™ Verification Module with factory-set values of 20.0 or 1.0 ±25%. Unique value
is assigned when dry verification is done immediately after calibration and is used as
pass/fail criteria for subsequent verifications.
Recommended
Maintenance
1.
Lamp replacement: once per year.
2.
Cleaning: mandatory before calibration, optional before calibration, and mandatory upon
verification failure.
Installation Environment
Indoor
Primary Compliance
Method
USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B
Limit of Detection
0.0032 NTU (according to criteria specified by ISO 15839)
1
All specifications are based on a calibration with 20.0 NTU formazin and with the offset turned off.
Table 1 1720E Low Range Specifications (continued)