3 event codes, Troubleshooting – Hach-Lange 1720E Low Range Turbidimeter User Manual

Page 35

Advertising
background image

31

Troubleshooting

7.3 Event Codes

Event codes are not displayed on controller and must be downloaded from the event log to
be viewed. Troubleshooting actions are provided in

Table 5 on page 30

.

High Readings

Deionized water turbidity is greater than
0.5 NTU

Clean the instrument.
Access Calibration History for turbidity value of
ultra-filtered water.
Verify the flow is between 200–750 mL/min.
Recalibrate the instrument.

Table 6 Additional Malfunctions Not Recorded in the Event Log (continued)

Symptom

Possible Cause

Corrective Action

Table 7 Event Log List

Event

Event #

Data1

Data2

Data3

Bubble Reject Change

0

0 = OFF
1= ON

Filter Size Change

1

0 = No avg, 1 = 6 sec
2 = 30 sec, 3 = 60 sec
4 = 90 sec

Data Log Interval Change

2

0 = 30 sec, 1 = 1min
2 = 5 min, 3 = 10 min
4 = 15 min

Power On

3

1

Calibration

4

Cal Gain

Initials

Verification

5

Expected Value

Meas Value

Initials

Dark Reading Warning

6

Measured Value

Temperature Warning

7

Temperature Value

Voltage Warning

8

16 = +5V high
32 = +5V low
64 = V in high
128 = V in low

Lamp Warning

9

1 = lamp V high
2 = lamp V low
4 = lamp I high
8 = lamp I low

A/D Fail

10

1

Lamp Fail

11

2

Flash Fail

12

4

Output Mode Change

13

0 = Active, 1 = Hold
2 = Transfer, 4 = Normal