Section 1 specifications – Hach-Lange 1720E Low Range Turbidimeter User Manual

Page 7

Advertising
background image

3

Section 1

Specifications

Specifications are subject to change without notice.

Table 1 1720E Low Range Specifications

Range

0–100 nephelometric turbidity units (NTU)

Measurement Units

mg/L, NTU, TE/F, FTU, Degree

Accuracy

1

± 2% of reading or ± 0.02 NTU (whichever is greater) from 0 to 40 NTU;
± 5% of reading from 40 to 100 NTU (when calibration is performed at 20.0 NTU with the
offset turned off).

Linearity

1

Better than 1% 0–40 NTU on formazin. Allows for accurate calibration at high turbidity values.

Resolution (Displayed)

0.0001 NTU

up to 9.9999 NTU; 0.001 NTU from 10.000 to 99.999 NTU; 0.01 NTU at 100.00

NTU

Repeatability

Better than ±1.0% of reading or ±0.002 NTU, whichever is greater

Response Time

For a full-scale step change, initial response in 1 minute, 15 seconds. Varies with flow rate,
see the table below. The response time is also dependent on the signal averaging time, which
is user selectable.

Sample Flow Required

250 to 750 mL/minute

Storage Temperature

–20 to 60 °C (–4 to 140 °F)

Operating Temperature

0 to 50 °C (32–122 °F) for single sensor system, 0 to 40 °C (32–104 °F) for two sensor system

Sample Temperature
Range

0 to 70 °C

Operating Humidity

5 to 95% non-condensing

Power Requirements

12 VDC ± 5%, 12.5 Watts maximum

Sample Inlet Fitting

1

/

8

barb fitting to ¼-inch NPT male adapter

Signal Average Time

no averaging, 6, 30, 60, and 90 seconds, user selectable. Default is 30 seconds.

Dimensions

Turbidimeter body and cap: 25.4 x 30.5 x 40.6 cm (10 x 12 x 16 inches)

Sensor Cable Length

2 m (6.6 ft); Optional 7.62 m (25 ft)

Mounting Options

Turbidimeter Body and Head Assembly: Wall; floor stand

Shipping Weight

1720E Series 2 Turbidimeter and Controller: 6.31 kg (13.5 lb); 1720E Turbidimeter only: 4.71
kg (10 lb)

Calibration Methods

1.

StablCal

®

(stabilized formazin) – primary or wet calibration of the instrument.

Recommended at 20.0 NTU.

2.

Formazin – user-prepared primary or wet calibration of the instrument. Recommended at
20.0 NTU.

3.

Multi-sensor calibration – Performed with a specialized calibration procedure for up to
eight sensors on a single set of fresh StablCal

®

standards.

% Step Change

Flow Rate

750

500

250

10

1¼ minutes

1½ minutes

2½ minutes

50

2 minutes

2½ minutes

6 minutes

90

3½ minutes

3½ minutes

9 minutes

99

4 minutes

5 minutes

12 minutes