Troubleshooting – Hach-Lange 1720E Low Range Turbidimeter User Manual

Page 34

Advertising
background image

30

Troubleshooting

Table 5

presents sensor warnings displayed in the Event Log, possible causes, and

corrective actions.

Table 6

presents additional malfunctions which may not be recorded in the Event Log.

Table 5 General Troubleshooting

Sensor Warning

Possible Cause

Corrective Action

Bad Lamp

Lamp burned out

Replace the lamp. See

section 6.4.3 on page 27

.

Lamp unplugged

Restore connection

+12 V connection loose at controller

Restore connection

Dislodged lamp

Reinstall lamp

Bad circuit board in turbidimeter head

Contact the Customer Service Department.

Low Signal

Photocell coated/dirty

See

section 6.4.1 on page 26

.

Contact the Customer Service Department.

Photocell wires disconnected

Reconnect wires

Photocell broken/cracked

Replace photocell
Contact the Customer Service Department.

Lens coated/dirty

Clean the lens using isopropyl alcohol and a cotton swab.

Obstructed light path

Remove obstruction

See Bad Lamp causes above

See Bad Lamp corrective actions above

Bad System Voltage

Improper wiring at controller

Refer to

section 3.1 on page 9

.

Turbidimeter head cable shortened to
improper length

Contact the Customer Service Department.

Fluctuation in voltage

Turn instrument power off and back on.

Bad circuit board in turbidimeter head

Contact the Service Department

A/D Converter Timeout

Fluctuation in voltage

Turn instrument power off and back on.

Bad circuit board in turbidimeter head

Contact the Customer Service Department.

High Dark Counts

Light Leak—Turbidimeter head not on
turbidimeter body or calibration
cylinder during Power Up or Zero
Electronics

Make sure the turbidimeter head is on the turbidimeter
body and properly aligned and repower instrument or
perform ZERO ELECTRONICS in the CALIBRATION
MENU.

Light Leak—Turbidimeter head not
properly aligned on the turbidimeter
body or calibration cylinder during
Power Up or
Zero Electronics

Make sure the turbidimeter head is properly aligned and
repower instrument or perform ZERO ELECTRONICS in
the CALIBRATION MENU.

Photocell broken/cracked

Contact the Customer Service Department.

Table 6 Additional Malfunctions Not Recorded in the Event Log

Symptom

Possible Cause

Corrective Action

Continuous Underrange The calibration standard was either

improperly prepared or was unstable at
the time the calibration was accepted.

Verify the accuracy of calibration standards and
recalibrate the instrument. See Low Signal in

Table 5

.

Continuous Overrange
(100 NTU)

The calibration standard was either
improperly prepared or was unstable at
the time the calibration was accepted.

Verify the accuracy of calibration standards and
recalibrate the instrument.

Erratic Readings

Inadequate bubble removal from sample Verify the accuracy of calibration standards and

recalibrate the instrument.
Increase the signal averaging time to a longer interval.
Make sure the Bubble Reject feature is turned on.
Slow the flow of sample into the instrument.