Hach-Lange 2100 AN Laboratory Turbidimeter User Manual User Manual

Page 13

Advertising
background image

Figure 5 Indicator lights

1 Lamp icon light

7 PRINT light

2 "CAL?" light

8 SIGNAL AVG light

3 "Manual" light

9 CAL/Zero light

4 "Auto" light

10 SETUP light

5 RATIO light

11 SAMPLE light

6 FLOW light

Table 2 Light descriptions

Light

Description

Illuminated when the instrument light source is on.
Flashes when there is not sufficient light for measurement.

CAL?

"CAL?" is shown during a calibration if the calibration data is not
within the acceptable range.
Flashes when the instrument should be calibrated.

Note: The CAL? light applies when the USEPA filter and a 25-mm sample cell
are used. Ignore the CAL? light if illuminated during calibration when a different
filter or a smaller sample cell is used. Push UNITS/Exit to start measurements.

Manual

"MANUAL" is shown above the Range Mode label when the
instrument is in manual ranging mode.

Auto

"AUTO" is shown below the Range Mode label when the instrument is
in auto ranging mode.

RATIO

Illuminated when Ratio is on.

FLOW

Illuminated when the Flow mode of operation is selected.
Flashes when the flow cycle is done.

PRINT

Illuminated when the printer interval feature is selected.
Flashes when a print interval has been selected but is not active.

SIGNAL

AVG

Illuminated when signal averaging is on.

CAL

Zero

Illuminated when Calibration or Zeroing mode is selected.

SETUP

Illuminated when Setup mode is selected.

SAMPLE Illuminated when Sample mode is selected.

English 11