Color measurement and calibration procedure, Color, Measurement and calibration procedure – Hach-Lange 2100 AN Laboratory Turbidimeter User Manual User Manual
Page 32
Advertising

Color measurement and calibration procedure
1. Make sure that the
455 nm filter assembly
is installed in the
instrument.
2. Push RANGE to
select manual or
automatic ranging.
Refer to
on page 32.
3. Push SIGNAL AVG
to set signal averaging
on or off. Refer to
4. Push UNITS/Exit
until "CU" is shown on
the display.
5. Push CAL/Zero.
The display shows
"ZERO" CU.
Note:
The instrument starts
analytical zeroing for
Color, Transmittance,
and Absorbance modes
at the same time.
6. Using the manual
flow cell kit, install the
flow cell. Refer to
on page 33.
Note: The sample cell
cover does not close
when the flow cell is
installed.