Hach-Lange 2100 AN Laboratory Turbidimeter User Manual User Manual

Page 51

Advertising
background image

• Replace the lamp with the same size, style and electrical rating

(4708900). Refer to

Replacement parts and accessories

on page 53.

• Do not touch the lamp as oil from skin will damage the lamp. Clean

the lamp with alcohol as necessary.

• Either lamp lead can be put in either terminal block position.

• Turn the instrument on 30 minutes (Ratio on) or 60 minutes (Ratio off)

before measurement or calibration.

• Calibrate the instrument after the lamp is replaced.

To remove the lamp, refer to the illustrated steps.
To install the lamp, do the illustrated steps in the opposite direction.