7. Read and record the
value when stable.
Note: To print or send
(via RS232) a
measurement record,
push PRINT.

Absorbance and transmittance measurement

Measurement notes

For the best accuracy and reproducibility:

• Set the zero reference point before measurement. Set the zero

reference point again when a measurement is not taken for several
hours as shown in

Absorbance and transmittance measurement

procedure

on page 28.

• Color, transmittance and absorbance measurements use the same

zero reference point. Color, absorbance and transmittance can be

measured on a single sample after setting a zero reference point in
one of the three modes.

• Use a flow cell for measurements. A flow cell is necessary to get the

accuracy and reproducibilty specifications shown in

Specifications

on page 5.
If a flow cell is not used, use a single indexed sample cell or match
sample cells. Sample cells should be matched using the
Transmittance or Absorbance modes. Refer to

Matching sample cells

on page 21.

• Refer to

Measurement notes

on page 25 for more measurement

notes.

English 27