Prepare the stablcal standards, Calibration notes, Prepare the stablcal standards calibration notes – Hach-Lange 2100 AN Laboratory Turbidimeter User Manual User Manual

Page 15

Advertising
background image

calibrations. The manufacturer cannot guarantee the performance of the instrument
if calibrated with co-polymer styrenedivinylbenzene beads or other suspensions.

Prepare the StablCal standards

When received and at intervals:

1. Clean the exterior surface of the StablCal vials with laboratory glass

cleaning detergent.

2. Rinse the vials with distilled or deionized water.
3. Dry the vials with a lint-free cloth.

Note: Never shake or invert the < 0.1 NTU standard. If the standard has been
mixed or shaken, do not move the vial for 15 minutes or more before using.
Note: Do not remove the caps from the sealed vials.

Make sure that the StablCal standards are at ambient instrument
temperature before use (and no greater than 40 °C (104 °F)).
Mix the standards before use:

1. Open the case lid. Remove the < 0.1 NTU standard from the plastic

case.

2. Leave the other standards in the case. Close the case lid.
3. Shake the case vigorously for at least 10 seconds.
4. Let the standards stand with no movement for 3–5 minutes before

use.

Calibration notes

• Make sure that the instrument is in the same ambient conditions as

where it is used.

• Make sure that the standards are at the same ambient temperature as

the instrument before use.

• Use only the provided silicone oil. This silicone oil has the same

refractive index as the vial glass and masks minor glass differences
and scratches.

• Store the oiling cloth in a plastic storage bag to keep the cloth clean.
• If power is lost during calibration, the new calibration data is lost and

the last calibration data is used. To exit a calibration and not save the
new values, push UNITS/Exit.

• In Calibration mode, automatic range and signal averaging on are

selected. When calibration is completed, all operational modes go
back to the last settings.

• All nephelometric (turbidity units of measure) calibrations are done at

the same time.

• Ratio-on and Ratio-off calibration data is measured and recorded at

the same time.

• The 7500-NTU standard does not have to be measured during

calibration if turbidity less than 4000 NTU will be measured. Push
CAL/Zero after the 4000 NTU standard is measured to complete the
calibration procedure.

• Clean the USEPA filter assembly before doing a primary calibration,

or at least every 3 months (which is the USEPA-recommended
primary calibration interval).

English 13