Section 4 operation, 1 software menu – Hach-Lange LT 200 Thermostat User Manual

Page 58

Advertising
background image

57

Section 4

Operation

4.1 Software Menu

SELECT BLOCK
LEFT/RIGHT

Select the heating block.

SELECT PROGRAM
100 °

Selection of 100 °C temperature program. Heats
cuvettes for 30, 60, or 120 minutes at 100 °C.

DIGESTION PERIOD?

120’/60’/30

Selection of temperature time.

100 °C: 100 °C 60’

OK

For example, the “100 °C, 60 minutes” setting is
needed to digest samples using the LCW 902
Crack Set. Start the program with

OK

.

SELECT PROGRAM
COD

Selection of COD Temperature program. Heats
cuvettes for 2 hours at 148 °C. In the cooling
phase, 4 beeps indicates that the cuvettes have
cooled to 80 °C. Remove and carefully invert the
cuvettes several times, before allowing them to
cool further in a rack.

COD: 148 °C 120’

OK

Start the program with

OK

.

SELECT PROGRAM
40 °

Selection of 40 °C temperature program. Heats
cuvettes for 10 minutes at 40 °C.

40 °C: 40 °C 10’

OK

Start the program with

OK

.

SELECT PROGRAM
PRG1

Programming a user-specific temperature program
(

5.4 User programs auf Seite 65

).

Advertising