Translation of the Original Operation Manual

Übersetzung der Originalbetriebsanleitung

Degaussing Overview

The LCD Display (3) gives status, automatic

count, total count, software version, field

strength, user instructions and fault messages.

The following information appears on the LCD

Display (3) from power-up through one automatic

degaussing cycle:

Automatic Degaussing Technical Detail

1) The LCD displays

“Insert Media Session =

0”.

2) The media to be degaussed is inserted into

the Media Drawer. While the drawer is open,

the LCD will display

“Drawer Open”.

3) Sensors in the degausser detect the open /

closed status of the drawer.

When the ERASE Push Button (6) is

depressed to initiate the degaussing process,

the degausser will confirm when the drawer

is fully closed and the erase cycle will begin.

The LCD will display

“Erase Cycle Initiated”

followed by

“Charging >>>>>”

4) When the capacitors are fully charged, the

capacitors energy will be released into the

degaussing coil to create the erasing field.

5) To ensure complete erasure, each erase cycle

is measured to ensure the strength of the

erase field is sufficient for complete erasure

of data. In addition, the LCD will display a

Field Strength Graph

“--------==I===” of the

actual erase field giving the operator real time

status of the degausser. The “

I” in the graph

represents the strength of the erase field. The

“=====” portion of the graph represents the

normal range for field strength level. The

“I”

within the

“--------=====” indicates the erase

field strength. The position of the

“I” can vary

from cycle to cycle as it is based on the actual

field strength.

6) If the field is in the normal range, the LCD will

display

“Data Eliminated”.

7) If the field is not in the normal range, the LCD

will display

“Erase Failure”. In this case, the

media should be erased again. To clear the

“Erase Failure” message, power cycle the

degausser by pushing the POWER Push

Button (4) off, wait one minute and then press

the power back on. If another erase failure

happens, call Products Customer Support for

assistance.

8) When the LCD displays

“Remove Media”.

Follow the steps to open the media drawer

and remove the degaussed media.

9) The degausser’s internal counter will advance

one count and display the total number of

media degaussed for the current session.

LCD DISPLAY

DESCRIPTION

Action / Step 1:

Press POWER Push Button (4) ON

Displays Manufacturer.

Displays Degausser Model.

Displays current firmware version.

Displays total erasing cycles. Number on

left indicates number of times the counter

reached 50,000.

Ready for media to be placed in the Me-

dia Drawer. Counter for erase cycles from

power up.

Indicates Media Drawer is open – this

Dis normal when inserting media to be

erased.

Action / Step 2:

Insert Media

Press ERASE Push Button (6)

Close Drawer (1)

Drawer has closed and the erase cycle has

started.

Displays capacitor charging status.

Visual indicator of erasing field strength.

Indicates power and erasing field strength

was verified good and data has been elimi-

nated.

Action / Step 3:

Remove Media

Erase cycle has completed. Open drawer

and remove media.