Laurel Electronics LAUREATE SERIES Serial Comunications User Manual
Page 33

6.3
WEIGHT METER
1-BYTE RAM DATA TABLE
Hex
Hex
Address Item Name
Value
6B
Configuration
Bit 7 6 5 4
3 2 1 0
0 = Linear Data
1 =Custom Curve (Extended DPM)
0 = Peak of net value
1 = Peak of gross value
0 = Setpoint off set enabled
1 = Setpoint offset disabled
0 = Setup scale method
1 = Reading 2 coord method
0 0 0 = Not Rate
0 0 1 = Rate X 0.1
0 1 0 = Rate X 1
0 1 1 = Rate X 10
1 0 0 = Rate X 100
1 0 1 = Rate X 1000
1 1 0 = Rate X 10,000
0 = Not used
BF
Analog Setup
Bit 7 6 5 4
3 2 1 0 bit
0 = Analog Out Unfiltered
1 = Analog Out Filtered
0 = Current Output
1 = Voltage Output
0 = Output of net value
1 = Output of gross value
0 0 0 0
0 = Not used
35
Decimal point
01
Byte values in hex
XXXXX.
02
(2 hex characters/byte)
XXXX.X
03
XXX.XX
04
XX.XXX
05
X.XXXX
06
.XXXXX
34
Lockout2
Bit 7 6 5 4
3 2 1 0 bit = 0 is unlocked for all items
1 = SEr 1, Ser 2, Ser3 locked
1 = An Lo, An Hi locked
1 = Alarm setpoint programming locked
1 = ALSEt locked
1 = Front panel DPM Reset locked
1 = Front panel Peak & Alarm Reset locked
1 = View Alarm setpoints locked
1 = View Peak value locked