Laurel Electronics LAUREATE SERIES Serial Comunications User Manual
Page 35

2F
Filter
Bit 7 6 5 4
3 2 1 0
0 0 0 0
Auto Filter
0 0 0 1
Batch (16 samples) filter
Time constant
60 Hz 50 Hz
0 0 1 0
Moving Average
.07s .085s
0 0 1 1
Moving Average
.14
.17
0 1 0 0
Moving Average
.28
.34
0 1 0 1
Moving Average
.57
.68
0 1 1 0
Moving Average 1.13
1.36
0 1 1 1
Moving Average 2.27
2.72
1 0 0 0
Moving Average 4.53
5.44
1 0 0 1
Moving Average 9.06 10.88
1 0 1 0
Unfiltered
1 = High adaptive threshold (0 = low threshold)
1 = Display filtered signa l (0 = display batch)
1 = Take peak of filtered signal (0 = unfiltered)
1 = Alarm from filtered signal (0 = unfiltered)
2E
Setup
Bit 7 6 5 4
3 2 1 0
EXT IN A
EXT IN B
0 0 0 0
Meter Reset
Meter Hold
0 0 0 1
Function Reset
Peak Display
0 0 1 0
Meter Hold
Peak Display
0 0 1 1
Meter Hold
Tare
0 1 0 0
Peak Display
Tare
0 1 0 1
Meter reset
Tare
0 1 1 0
Function reset
Tare
0 1 1 1
Tare reset
Tare
1 0 0 0
Display blank
Tare
1 0 0 1
Meter reset
Display blank
1 0 1 0
Function Reset
Display blank
1 0 1 1
Display item
Tare
1 1 0 0
Display item
Display blank
1 1 0 1
Meter reset
Display item
1 1 1 0
Function reset
Display item
1 1 1 1
Meter hold
Display item
0
Scale using scale and offset
1
Scale using coordinates of two points
0 Peak key displays peak
1 Peak key tares net value
0 No dummy zero
1 Displays dummy zero
-34-