Laurel Electronics LAUREATE SERIES Serial Comunications User Manual
Page 36

3-BYTE RAM MEMORY DATA
All items except Scale Factor use this format:
Note: Hex value (2's complement)
MS byte
Mid byte
LS byte
X X
X X
X X
Scale factor uses this format:
MS byte
Mid byte
LS byte
MS Bit ( 0 = Polarity and Decimal Point)
0 X
X X
X X
Positive
1 Negative
9
Decimal point
XXXXX.
2
A
Decimal point
XXXX.X
3
B
Decimal point
XXX.XX
4
C
Decimal point
XX.XXX
5
D
Decimal point
X.XXXX
Note: Hex value (Absolute value)
6
E
Decimal point
.XXXXX
2C
Alarm Cnfg2
Bit 7 6 5 4
3 2 1 0
Alarm Trigger Delay
60 Hz
50Hz
0 0 0
.018s
.021s
0 0 1
.035
.043
0 1 0
.07
.085
0 1 1
.14
.17
1 0 0
.28
.34
1 0 1
.56
.68
1 1 0
1.13
1.36
1 1 1
2.27
2.72
Alarm comparison
0 0 0 0
0 AL1 Net, AL2 Net
0 0 0 0
1 AL1 Gross, AL2 Net
0 0 0 1
0 AL1 Net, AL2 Gross
0 0 0 1
1 AL1 Gross, AL2 Gross
2B
Alarm Cnfg1
Bit 7 6 5 4
3 2 1 0
0 0 0 0
AL1 hi activ, AL2 hi activ
0 0 0 1
AL1 lo activ, AL2 hi activ
0 0 1 0
AL1 disabled, AL2 hi activ
0 1 0 0
AL1 hi activ, AL2 lo activ
0 1 0 1
AL1 lo activ, AL2 lo activ
0 1 1 0
AL1 disabled, AL2 lo activ
1 0 0 0
AL1 hi activ, AL2 disabled
1 0 0 1
AL1 lo activ, AL2 disabled
1 0 1 0
AL1 disabled, AL2 disabled
0 0
AL1 non-latch, AL2 non-latch
0 1
AL1 latch, AL2 non-latch
1 0
AL1 non-latch, AL2 latch
1 1
AL1 latch, AL2 latch
0 0 Relay 1 ON, Relay 2 ON
0 1 Relay 1 OFF, Relay 2 ON
1 0 Relay 1 ON, Relay 2 OFF
1 1 Relay 1 OFF, Relay 2 OFF