Luminex 200 analyzer specifications – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 29

Advertising
background image

x

MAP Technology

System Overview

PN 89-00002-00-150 Rev. A

3 - 3

Sensitivity

•

Detect 1000 fluorochromes phycoerythrin (PE) per xMAP
microsphere

•

Reporter channel dynamic range: 3.5 decades of detection

Capacity

The specifications below reflect minimum capacity values:

•

Distinguish a minimum of 1 to a maximum of 100 unique xMAP
microsphere sets in a single sample

•

Detect and distinguish surface reporter fluorescence emissions at
575 nm on the surface of 1-100 unique xMAP microspheres sets
in a single sample

•

Sample core: 15-20 µm core at 1 µL/sec. sample inject rate

•

Maintain samples at a constant temperature from 35°C to 55°C
(95°F to 131°F)

•

Automatic sampling from a 96-well plate

•

Start sampling from any well position

•

Sheath container and waste container hold enough volume to run
up to two 96-well plates between refills

•

Microtiter plates with 96 wells must be compatible with the
Luminex XYP instrument plate holder. The following microtiter
plate types are compatible with the Luminex XYP instrument
plate holder: any color of flatbottom, conical, round, filter
bottom, or half-well plates. The overall plate must not have an
overall height of more than 0.75” (19 mm).

•

Microtiter plates with 96 wells must be compatible with
Luminex XYP instrument heater block temperature from 35°C to
55°C (95°F to 131°F) when performing heated assays and using
the heater block.

Luminex 200

Analyzer

Specifications

•

Indoor use only

•

Operating temperature: 15°C to 30°C (59°F to 86°F)

•

Humidity: 20% to 80%, noncondensing

•

Altitude: Operation up to 2400 m (7874 ft.) above mean sea level

•

Physical dimensions: 43 cm (17 inches) W x 50.5 cm (20 inches)
D x 24.5 cm (9.5 inches) H

•

Weight: maximum of 25 kg (60 lbs.)

•

UL installation category: UL Installation Category II, as defined
in Annex J of UL 61010A-1

•

Pollution degree: UL Pollution Degree 2, as defined in Section
3.7.3.2 of UL 61010A-1

•

Shipping and storage: The allowable shipping and storage
temperature and humidity ranges are 0°C to + 50°C and 20-80%
noncondensing, respectively

•

Input voltage range: 100 - 120 V~ ± 10%, 1.4 Amp, and 200-240
V~ ± 10%, 0.8 Amp, 47-63 Hz.