The message at the top reflects the current status of the Luminex
analyzer.

Ready: Indicates that the Luminex analyzer is ready to collect data.

Pressure: Pressure normally displays during a routine pressure
check at the beginning of each run. However, if the pressure message
remains for more than 1.5 minutes, a lack of pressure exists in the
system. Check for the following problems:

•

Low sheath fluid

•

Loose hose connections

•

Loose cap on sheath fluid bottle.

•

Leak in sheath fluid container, tubing, or bubble trap

Not Ready: Normally indicates that the analyzer is initializing.

Busy: Indicates an intermediate state between Standby and Ready.

Standby: Shows that the flow of sheath fluid stopped; the system is
idle.

Bad Link: The communication between the analyzer and the
computer is not working. Check that:

•

The analyzer is powered on.

•

The cables between the analyzer communications port and
computer are properly installed.

If the situation has been corrected, restart the Luminex software
application to reinitialize this link.

Warm Up: Indicates a 30-minute warmup of the analyzer’s optics.
While powered, the optics remain active for four hours after each
acquisition.

Running: Indicates that the Luminex analyzer is collecting data.

Sample empty: Indicates an empty sample syringe. The system
automatically cancels data collection and adds a comment in the
notes section.

Connect: Indicates that the communication between the analyzer
and the computer was broken, then re-established. You must restart
the Luminex software application to reinitialize communication
links.

Sheath fluid: Indicates empty sheath fluid. Fill the sheath fluid
container before continuing.