Micromod MOD: 30ML Operation and Template Setup User Manual

Page 61

Advertising
background image

MOD 30ML Multiloop Controller

OPERATION

3-29

3.15 STARTUP

Procedures may vary from application to application, depending upon individual requirements
and control loop setup. The initial operating mode (usually manual), the set-point source
(usually local), set-point value, and control output can be configured for specified safe values
when power is first applied to the instrument. Each time a control application is first started
up, check all these items before turn-on.

3.15.1 Startup Without Set-Point Tracking

To start up a control loop in the manual mode:

1. Select the required loop using the TAG key.

2. Place the loop in manual mode with local set-point.

3. Adjust local set-point to starting value required for process variable.

4. Adjust manual output to a safe starting value for the process. This is usually below 0% or

above 100% of span.


5. Open block valve ahead of control valve or perform any other procedure required to

make the process variable available, such as starting a pump, turning on a power supply,
etc.


6. Manually adjust control output to bring process variable to set-point value. Ramping the

output should be slow enough to allow the process to respond to the change without
overshooting the set-point. Process is now operating under manual control with process
at the set-point.


3.15.2

Startup With Set-Point Tracking

The startup procedure is simplified when set-point tracking is used. With the control loop in
the manual mode, the set-point tracks the process so that the process and set-point values
are together when the operator switches to the automatic mode.

To start up the controller with set-point tracking:

1. Select the required loop using the TAG key.

2. Place the loop in manual mode.

3. Adjust manual output to a safe starting value for the process. This is usually below 0% or

above 100% of span.


4. Open block valve ahead of control valve or perform any other procedure required to

make the process variable available, such as starting a pump, turning on a power supply,
etc.

Advertising