STG3008-FA Manual

22

In the following example, the stimuli on output channels 1, 2, 4, 5 and channel 7 and 8,
that are all triggered by Trigger 1, are repeated five times after the trigger event, each time
accompanied by the Sync Out 1 output.

Repeating complete pulse protocols, Gating Trigger

Repeated stimulation of complete protocols can also be controlled by a gating trigger. Channels
associated with the Trigger input will be active as long as the condition of the Trigger is 1 (HIGH)
and switched off as soon as the condition becomes 0 (LOW). Gating trigger function can be
enabled in the Trigger Settings menu.