12. Error Codes

6.4 Select the sterilization program

Select the appropriate sterilization temperature

and cycle. (For additional information, see
appendix 2.)

6.5 Running the sterilization program

After selecting the appropriate sterilization

settings, load the samples onto the loading tray
and insert the tray into the autoclave using the
tray handling tool.

Instructions manual

Instructions manual

Code

E1

E2

E3

E4

E5

E6

E7

E8

E9

EE

EF

Description

Proposed solution

Steam generator temperature
sensor error

Temperature sensor of chamber

wall error

Inner temperature sensor error

Failure to raise temperature

Failure to release steam

Door has opened during the cycle

Time Error

Steam generator overheat

Failure to initialize

Failure to hold temperature

Program manually interrupted

Check steam generator temperature
sensor

Check inner temperature sensor

Check temperature sensor of chamber
wall

Check the water pump or the door seal

Check the air release valve

Make sure you have turned the door handle
to the max. position or check the door switch

Check the water pump and the
air release valve

Check steam generator temperature
sensor

Shut off the power and restart the
machine

Shut off the power and restart the
machine

Ensure that the clean water tank is not
empty or contact your dealer to check the
heating system and temperature sensors.

* Verify all samples are not sealed, capped or closed.

* Don't overload the trays above the stated limit (see Appendix 1).

* Don't stack the trays one above the other or put them in direct contact with

the walls of the sterilization chamber.

* When handling the trays, always use the tray removal tool.

* Wrap each sample separately, if samples are wrapped together verify that

they are of the same material.

* Seal the wrap with sterilization adhesive ribbon or by a thermal sealer.

* Never seal with metallic clips, pins, etc. as this jeopardizes the maintenance

of the sterility;

* When using sterilization paper, set the plastic part downward (tray side)

and the paper side facing upward.

Always wrap samples.