Installation, Appendix 2, Diagrams of the sterilization programmes – Nova-Tech B4000-16 User Manual

Page 7: Above dimensions are shown in mm

Advertising
background image

1 9

4

4. Installation

DIAGRAMS OF THE STERILIZATION PROGRAMMES

APPENDIX 2

PROGRAM

Temperature

(°C)

Pressure

(PSI)

Holding

time

(min.)

To

tal

time

(min.)

TYPE

MAXLOAD

(kg)

UNWRAPPED

134

121

134

121

30

16

30

16

4

15

6

15

12~30

30~45

30~40

35~50

WRAPPED

U n w r a p p e d s o l i d

m a t e r i a l

Unwrapped hollow

material

5.00

5.00

5.00

4.00

<100

0

15

DRY ONLY

Single-wrapped solid

material

Instructions manual

Instructions manual

LIQUID

121

134

30

16

12

30

30~65

45~65

Wrapped liquid

Wrapped liquid

0.50

0.50

*

*

*

*

*

Ensure that the sterilizer is installed with 2.5in. (10cm) ventilation space

on all sides of the sterilizer, and 5 in. (20cm) on top side. The clearance

required to open the door is 15.5in. (40cm).

The sterilizer should be placed on a level worktable.

Do not cover or block the door, ventilation or radiation openings on the

sterilizer.

Do not install the sterilizer near a sink or in a location where it is likely to

be splashed.

Do not install the sterilizer nearby a heat source.

Above dimensions are shown in mm.

The max. temperature of the 134 C sterilization cycle is 136C
The max. temperature of the 121 C sterilization cycle is 123C