Draining the water tanks, 5 replacement of the door seal – Nova-Tech B4000-16 User Manual

Page 12

Advertising
background image

1 4

9

10.5 Replacement of the door seal

Instructions manual

Instructions manual

7. Draining the water tanks

1.

Connect the supplied drain hose to

the drain valve by pressing it on firmly.

2.

Set the drain valve to the open

position by turning it counter clockwise

3.

Pull the drain valve outward, the

tank will begin to drain through the hose.

4.

After you have completed draining the

water, push the drain valve inward and
set to the closed position

The sterilizer includes two draining valves. The left valve is used
for draining the used water tank. The right valve is used for draining
the clean water tank. Always be sure that the hose is draining into
a vessel with at least 2L capacity.

10.5.1 Fully open the door.

10.5.2 Remove the door seal carefully by pulling it by hand away from

the door

10.5.3 Clean the new door seal carefully with a soft cloth saturated

with distilled water.

10.5.4 Moisten the new seal with medical disinfectant .

10.5.5 Insert the new seal following the instruction below:

1). Press on the top of the seal.

2). Press on the bottom of the

seal.

3). Press on the left and right

portions of the seal.

4). Press the remaining seal surface

and ensure that it is smooth.

Caution: Always ensure the chamber and the door have
been cooled down before attempting to change the seal.

Advertising