8 Advanced adjustment settings

8.1 Adjustment of sterilization time and drying time

8.1.1 Hold the UNWRAPPED button for 30seconds. The machine will beep

to signify that you have enetered the advanced settings mode.

8.1.2 The pressure display window shows “0” and the temperature display

window shows - - -. The timer display windowwill show the pre-
programmes sterilization time parameter.

8.1.3 Select the sterilization program that you wish to adjust. For example,

If you want to adjust the parameter of

Press the

button. Press the button again ,the time display shows “1”.

WRAPPED.

WRAPPED

,

0 means the 121 sterilization time.(The range from 3 to 40 min.)

1 means the 134 sterilization time.(The range from 3 to 40 min.)

2 means the drying time.(The range from 3 to 20 min.)

8.1.4 Press

button, the value of time parameter will plus one.

8.1.5 After finishing the adjustment, press

button to confirm it.

8.1.6 When adjusting the

program, there is only one option,

2. This is to adjust the drying time.

TEMP

START

DRYING

Caution: Always ensure that the sample being sterilized can
withstand the sterilization time that is being programmed.

Instructions manual

Instructions manual

10.3 Cleaning the inner-chamber, weekly

-

Remove the trays and tray rack from the inner-chamber

- Wipe the inner chamber with a soft cloth saturated in distilled water
- Apply the same procedure to the trays and tray rack.

Fig. 2

Fig. 1

Caution:
Never attempt to adjust the door seal when the door is in the
closed position.

10.4 Door adjustment

Below are the instructions for the door adjustment. This should only be
performed if the door is not providing the proper seal or if the door seal has
just been replaced.

10.4.1 Open the door and insert the door adjustment tool in the gap beneath the

plastic cover (Fig. 2).

10.4.2 To tighten the door seal, use this tool to adjust the door adjustment nut

counter clockwise (Fig. 1).

10.4.3 If the door seal is too tight (it is difficult to lock/unlock the door),

you may also adjust the door adjustment nut clockwise to loosen it.