5 manual calibration – Nova-Tech SALT 6+ User Manual

Page 18

Advertising
background image

Instruction Manual

COND 6+, TDS 6+, SALT 6+

-13-

3.5 Manual

Calibration

In Manual Calibration mode you are not limited to the conductivity calibration standards
listed previously in Table 1. This example shows a manual calibration sequence using a
12.00 mS/cm conductivity calibration standard.

1.

If necessary, press MODE key to select conductivity
mode.

1.

Rinse the probe thoroughly with de-ionized water or
a rinse solution, then rinse with a small amount of
calibration standard.

1.

Dip the probe into the calibration standard. Stir the
probe gently to create a homogeneous sample.
Allow time for the reading to stabilize.

1. Press

CAL

to enter conductivity calibration mode.

The [CA] indicator will appear briefly, then a value
will appear flashing.

NOTE: To exit calibration without confirmation, press

CAL to return to measurement mode.

2.

When the value is stable, press ▲ or ▼ as needed
to adjust the value to match your calibration
standard.

3. Press

ENTER

to confirm the adjusted value. [CO]

will appear briefly indicating that the calibration was
successful. The meter returns to measurement
mode.

4.

Repeat steps 1-6 as needed with additional
calibration standards.

mS

mS

mS

mS

CAL

HOLD

ENTER

Measurement Mode

Advertising
This manual is related to the following products: