2 select cell constant (cond 6+, tds 6+ only), 3 automatic calibration (cond 6+ only) – Nova-Tech SALT 6+ User Manual

Page 25

Advertising
background image

Instruction Manual

COND 6+, TDS 6+, SALT 6+

-20-

5.2

Select Cell Constant (COND 6+, TDS 6+ only)

Your meter includes a probe with a nominal cell constant (k) of 1.0. Use probes with k =
0.1 and 10 (sold separately) for improved performance in extreme samples. Use this
setup function to change the cell constant if necessary. Meter default is 1.0 to match the
included probe.

k = 0.1 ideal for low measurements <20 µS/cm (<10 ppm).

k = 1.0 ideal for mid-range measurements

k = 10 ideal for high measurements >20 mS/cm (>10 ppt).

1.

Enter advanced setup as described in Section 5.1.

1. Press

▲ or ▼ until [CELC] appears. Press ENTER.

2. Press

▲ or ▼ to select “1.0”, “0.1” or “10.0”. Ensure that

the cell constant corresponds with the electrode you are
using.

3. Press

ENTER

to select and return to [CELC] setup

function.

4. Press

▲ or ▼ to move to the next setup function or

press CAL to exit to measurement mode.

5.3

Automatic Calibration (COND 6+ only)

Select automatic calibration “YES” for easy calibration of (4) factory pre-set conductivity
calibration standards (see Section 3.3 Table 1). To use other standards or to calibrate
any standard manually select “no”. Default value is “YES”.

5.

Enter advanced setup as described in Section 5.1.

6. Press

▲ or ▼ until [ACAL] appears. Press ENTER.

7. Press

▲ or ▼ to select [YES] or [no].

8. Press

ENTER

to select and return to [ACAL] setup

function.

9. Press

▲ or ▼ to move to the next setup function or

press CAL to exit to measurement mode.