Instruction Manual

COND 6+, TDS 6+, SALT 6+

-19-

Entering Advanced Setup Mode

Cell Constant. Select k = 0.1, 1.0, or 10.0.

Default value is 1.0. (COND 6+, TDS 6+ only)

Select Automatic Calibration. “Yes” for auto calibration
and “No” for manual calibration.

Default value is “Yes”. (COND 6+ only)

%

Adjust Temperature Coefficient from 0.0 to 3.0 % per °C.

Default value is 2.1 % per °C.

C

°

Normalization Temperature. Select “20 °C” or “25 °C”.

Default value is 25 °C.

Adjust TDS factor from 0.4 to 1.0.

Default value is 0.5. (TDS 6+ only)

Select Single Point Calibration. Select “Yes” or “No”.

Default value is “Yes”.

Select Measurement Unit, TDS (ppt) or percentage (%).
(SALT 6+ only)

User reset to factory defaults. Select “Yes” or “No”.

Default value is “no”.

Overview of Advanced Setup