5.8

Restore Factory Default Values

Use this function to reset all parameters to factory default settings. This clears all
calibration data and any other setup functions you might have
changed.

IMPORTANT:

Once activated the settings and calibration data

will be erased and can not be undone.

1.

Enter advanced setup as described in Section 5.1.

2. Press

▲ or ▼ key until [UrSt] appears. Press ENTER.

3. Press

▲ or ▼ to select [Yes] or [no].

4. Press

ENTER

to confirm.

5.

If reset [Yes] confirmed, the meter will return to
measurement mode after initialization.

6.

PROBE CARE AND MAINTENANCE

Keep the probe clean. Rinse the probe twice, and gently swirl
it while you take readings. For best results, soak a dry probe
for at least 5-10 minutes before calibration. Rinse the probe
with clean water before storing. Never scratch the bands with
an abrasive or hard substance. Do not strike against hard
surfaces or submerge the cable for extended periods.

Do not immerse the probe in oily solutions, aggressive
solvents, or strong acids. Clean with a mild detergent or
isopropyl alcohol then rinse thoroughly. Dry storage is
acceptable. Recalibrate after cleaning.

The included conductivity probe (EC-CONSEN91B) features
a removable probe guard for easy cleaning. To remove —
grip yellow probe guard and twist clockwise
to release locking notch, then slide off.

NOTE:

Remember to re-attach the probe guard – failure to

do so will result in erroneous readings!