6 temperature calibration – Nova-Tech SALT 6+ User Manual

Page 19

Advertising
background image

Instruction Manual

COND 6+, TDS 6+, SALT 6+

-14-

3.6 Temperature

Calibration

The electrode includes a built-in temperature sensor that
is factory calibrated with the meter. Perform temperature
calibration only if you suspect temperature errors may
have occurred over time or when the probe is replaced.

You can offset the temperature reading up to ±5 °C from
the original (default) reading.

1.

Connect the mini phono plug of the electrode to the
meter. See Section 3.7.

2.

If necessary, press MODE to select temperature
measurement mode.

3. Press

CAL

to initiate temperature calibration. “CA

will appear briefly then a temperature value will start
flashing.

4.

Dip the probe into a solution with known
temperature (for example, a temperature bath).
Allow time for the temperature to stabilize.

5.

When the value is stable, press ▲ or ▼ as needed
to adjust the value to the solution temperature.

6. Press

ENTER

to confirm the adjusted value. [CO]

will appear briefly indicating that the calibration was
successful. The meter returns to measurement
mode.

NOTE: To exit calibration without confirmation, press

CAL to return to measurement mode.

C

°

C

°

C

°

HOLD

ENTER

° C

CAL

Advertising
This manual is related to the following products: