Sensor usage tips, Calibrating the temperature probe, Suggested activities – PASCO PS-2147 High Resolution pH-ORP-ISE Amplifier User Manual

Page 4

Advertising
background image

CAUTION: Do not place the PS-2147 pH/ORP/ISE Sensor case in liquids

or expose to chemicals. The connectors and internal electronic circuit may be
damaged.

012-08636B

Sensor Usage Tips

PASCO Equipment used with the
pH/ORP/ISE with Temperature Sensor

• Compare ORP of water from a faucet, fish tank, and a pond.
• Discover how long chlorinated tap water needs to stand before adding to a

fish tank. Analyze the effect of temperature on this process.

• Study the effects of temperature on pH in a buffer
• Observe the heat of neutralization in a titration
• Measure the freezing and boiling point of a solution
• Study the effects of temperature and dilution on buffers

CARD 2B

Calibrating the Temperature Probe

Calibration of the temperature probe is not usually required. The accuracy of the
PASCO temperature probes is better than 0.5

°

C across the entire range of the

probe. However, the probe may be calibrated using the following procedure.

1. Plug the PS-2147 Sensor into a PASPORT interface connected to a USB-

compatible computer.

2. Insert a temperature probe into the sensor box using the 3.5

mm input on

the right side of the sensor box.

3. Click the Setup button in the DataStudio toolbar. The Setup window will

open.

4. Click on the Temperature Calibrate button in the Setup window. The

Calibrate window will open.

5. Place the temperature probe in an ice water bath.
6. Ensure the value in the dialogue box for Point 1 is 0. This value may be

changed to the temperature of the controlled temperature bath if it is
different than 0

°

C.

7. Click the Set button adjacent to the Point 1 dialogue box.
8. Place the temperature probe in boiling water.
9. Ensure the value in the dialogue box for Point 2 is 100. This value may

be changed to the temperature of a controlled temperature bath if it is
different than 100

°

C.

10. Click the Set button adjacent to the Point 2 dialogue box.
11. To save the calibration values, click the OK button. The calibration is

complete.

Temperature Probes
• PS-2135 Fast Response

Temperature Probe (3-pack)

• PS-2131 Skin/Surface

Temperature Probe

• PS-2153 Stainless Steel

Temperature Probe

Related Sensors
• PS-2117 Drop Counter

Ion Selective Electrodes
• CI-6717 Ammonium
• CI-6727 Calcium
• CI-6726 Carbon Dioxide (aqueous)
• CI-6732 Chloride
• CI-6728 Fluoride
• CI-6736 Lead
• CI-6735 Nitrate
• CI-6716 Oxidation-Reduction Potential (ORP)
• CI-6733 Potassium
• CI-6734 Sodium

Suggested Activities