Test Method Description
The total hardness ezSample test method
employs ethylenediaminetetraacetic acid (EDTA) titrimetric
chemistry.
1,2
In an alkaline solution, EDTA forms a chelated soluble complex with calcium and
magnesium ions. Calmagite is used as the endpoint indicator. Results are expressed as calcium
carbonate (CaCO
3
).
1,2,3
NOTE: Because the ampoules have nonlinear scales, the accuracy of the ezSample field titrator kit varies with the
analyte concentration. At the low end of the test range, the accuracy is
± 5%. At the high end of the range, the
accuracy falls to
± 20%.