Smart Timer

012–06734A

10

Pulley (rad/s): The Smart Timer will measure the speed of a pulley passing through a

photogate in units of radians/second. One measurement will be taken each time the Start/

Stop switch is pressed. The Smart Timer cannot differentiate between clockwise and

counterclockwise directions. Note that, as in many other modes, if a “*” shows in the first

character position of the second line, the Smart Timer is actively waiting for an external

timing event to occur. If the “*” is not showing, the Timer measurement mode may be

changed. This mode has a minimum speed requirement. The photogate must be blocked

twice within two seconds to obtain a valid reading, translating to a minimum speed of

0.31 rad/s.

Pulley (rev/s): Besides displaying in different units, this mode uses the display differently

than in the radians/second measurement. The display provides a real-time measurement of

the speed of the pulley by updating once per second. Once a speed is displayed, press the

Start/Stop key to freeze the display. The Smart Timer indicates that the measurement is

frozen by displaying a “!” in the first column. Press Start/Stop to erase the “!” and start

collecting new measurements. Any time the first column is blank or has a “!” displayed,

the type of measurement being done can be changed by pressing the Select

Measurement or Select Mode keys. To move to a different measurement, press Start/

Stop again. Like the radians/second measurement, there is a two-second maximum time

for two photogate blocks, translating to a minimum speed of 0.05 rev/s. Also note that

accuracy decreases rapidly as pulley speeds increase above 600 rpm. At 600 rpm,

accuracy is 1%.

Acceleration Modes

One Gate: The Smart Timer uses the time measurement between two equally spaced (5

cm) block/unblock/block sequences to calculate average acceleration. In activities using

the Smart Timer Picket Fence, the Picket Fence must be positioned so that the photogate

blocks only the 3-bar segment (5cm fence) of the Picket Fence. Note that the Smart Timer

is able to discern between acceleration (positive number) and deceleration (negative

number).

Linear Pulley: In this mode, the Smart Timer converts rotary motion of a PASCO pulley

to an equivalent linear acceleration in cm/s

2

.

Angular Pulley: In this mode, the Smart Timer converts rotary motion of a PASCO

pulley to an equivalent angular acceleration in rad/s

2

.

Two Gates: When two photogates are placed an arbitrary distance apart, the average

acceleration between the two can be calculated. In this mode, the inputs used are not

arbitrary. The first photogate to be encountered must be connected to input #1 and the

second to input #2.

Count Modes:

30 Seconds: The timed 30-second count mode will count high-to-low voltage transitions

on either input and display them on the second line of the liquid crystal display. After the

counting period is over, the Smart Timer will beep once, remove the power to the external

device, and freeze the display. Pressing Start/Stop erases the old count and begins a new