Operational theory – Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual

Page 19

Advertising
background image

Trace Oxygen Analyzer

Operational Theory

Teledyne

Analytical

Instruments

7

Operational Theory

2.1 Introduction

The analyzer is composed of three subsystems:

1. The zirconium oxide (ZrO

2

) sensor with a 0-25%

analysis range

2. Sample System
3. Electronic Signal Processing, Display and Control

The sample system is designed to accept the sample gas and

transport it through the analyzer without contaminating or altering the
sample prior to analysis. The ZrO

2

sensor is a device that translates the

amount of oxygen present in the sample into a voltage output. The
electronic signal processing, display and control subsystem simplifies
operation of the analyzer and accurately processes the sampled data. The
microprocessor controls all signal processing, input/output and display
functions for the analyzer.

2.2 ZrO

2

Sensor

2.2.1 Principles of Operation

The zirconium oxide (ZrO

2

) sensor is a solid state sensor. It is a

voltage device with a logarithmic response. The sensor uses a stabilized
zirconia disc as an electrolyte with a sensing electrode (the cathode) and
a counter electrode (the anode) on each side of the disc and contains a
thermocouple. The sensor is designed to have a small sample of the
unknown gas passed inside the sensor tube. Air (20.95% oxygen)
surrounds the outside of the tube. A heater is mounted around the sensor
to keep the sensor hot. The sensor construction is shown in Figure 2.1.

CAUTION:

THE SENSOR IS HOT. DO NOT TOUCH THE SENSOR
OR ALLOW FLAMMABLE MATERIAL TO COME IN
CONTACT WITH THE ASSEMBLY.

Advertising