Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual
Page 63

Trace Oxygen Analyzer
Operation
Teledyne
Analytical
Instruments
51
ppm O
2
Voltage Signal
Output (VDC)
Current Signal
Output (mA dc)
0 0.0
4.0
10 0.1
5.6
20 0.2
7.2
30 0.3
8.8
40 0.4 10.4
50 0.5 12.0
60 0.6 13.6
70 0.7 15.2
80 0.8 16.8
90 0.9 18.4
100 1.0 20.0
The analog output signal has a voltage which depends on the
oxygen concentration AND the currently activated analysis range. To
relate the signal output to the actual concentration, it is necessary to
know what range the instrument is currently on, especially when the
analyzer is in the auto ranging mode.
To provide an indication of the range, a second pair of analog
output terminals is used. They generate a steady preset voltage (or
current when using the current outputs) to represent a particular range.
The following table gives the range ID output for each analysis range:
Range
Voltage (V) Current (mA)
LO 0.25
8
MED 0.50
12
HI 0.75
16
CAL (0-25%)
1.00
20