Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual

Page 34

Advertising
background image

Installation 3000ZA2G


Teledyne

Analytical

Instruments

22

Span: Floating input. 5 to 24 V input across the + and – pins puts the

analyzer into the Span mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Span

mode to terminate when done. A synchronous signal

must open and close external span valve appropriately. See
Figure 3-5 Remote Probe Connector. (The –C option internal
valves operate automatically.)

Cal Contact:

This relay contact is closed while analyzer is spanning

and/or zeroing. (See Remote Calibration Protocol below).

Table 3-3: Remote Calibration Connections

Pin Function

9

+ Remote Zero

11

– Remote Zero

10

+ Remote Span

12

– Remote Span

40 Cal

Contact

41 Cal

Contact

Remote Calibration Protocol:

To properly time the Digital Remote Cal

Inputs to the Model 3000ZA2G Analyzer, the customer's controller must
monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote

Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It

will ignore your request to calibrate, and it will not remember that
request.

Once a zero or span command is sent, and acknowledged (contact

closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.

For example:

1. Test the CRC. When the CRC is open, send a zero command

until the CRC closes (The CRC will quickly close.)

2. When the CRC closes, remove the zero command.

Advertising