7 the analyze function, 8 signal output – Teledyne 3000ZA2G - Zirconium oxide analyzer User Manual
Page 62

Operation 3000ZA2G
Teledyne
Analytical
Instruments
50
L—10
M—1000
H—250000
Mode—FX/MED
or
L—10
M—1000
H—250000
Mode—FX/ HI
Press Escape to re-enter the Analyze mode using the fixed range.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit (or default value)
as established by the operator for that particular range, the
output saturates at 1 VDC (or 20 mA). However, the digital
readout and the RS-232 output continue to read the true
value of the oxygen concentration regardless of the analog
output range.
4.7 The Analyze Function
Normally, all of the functions automatically switch back to the
Analyze
function when they have completed their assigned operations.
Pressing the Escape button in many cases also switches the analyzer
back to the Analyze function. Alternatively, you can press the Analyze
button at any time to return to analyzing your sample.
4.8 Signal Output
The standard Model 3000ZA2G Trace Oxygen Analyzer is
equipped with two 0–1 VDC analog output terminals accessible on the
back panel (one concentration and one range ID), and two isolated 4–20
mA DC current outputs (one concentration and one range ID).
See Rear Panel in Chapter 3, Installation, for illustration.
The signal output for concentration is linear over the currently
selected analysis range. For example, if the analyzer is set on range that
was defined as 0–100 ppm O
2
, then the output would be: