Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

21

System

Alarm:

Actuates when DC power supplied to circuits is unacceptable
in one or more parameters. Permanently configured as
failsafe and latching. Cannot be defeated. Actuates if self test
fails.

(Reset by pressing button to remove power to relay. Then
press again and any other button Up and Down
simultaneously to resume.

Further detail can be found in Chapter 4, Section 4.6.

Table 3-2: Alarm Relay Contact Pins

Pin Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact

Digital Remote Cal Inputs:

Accept 0 V (off) or 24

VDC (on) inputs for

remote control of calibration. (See Remote Calibration Protocol below).
See Table 3-3 for pin connections.
Zero: Floating input. 5 to 24 V input across the + and – pins puts the

analyzer into the Zero mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Zero

mode to terminate when done. A synchronous signal must

open and close the external zero valve appropriately. See
Remote Probe Connector. (The –C option internal valves
operate automatically.)