Appendix Model

7600

Teledyne Analytical Instruments

116

CO concentration peak count alarm: (added only for CO/O

2

measurement) Alarm output turns on
according to the pre-set concentration and
count.

Whenever the instantaneous value of CO

exceeds the preset concentration value,
count increments. If the count exceeds the
preset value in one hour, the alarm
contacts close.

Communication function: RS-232C (9 pin D-sub)

Half-duplex bit serial
Start-stop synchronization
Modbus TM protocol

Contents:

Read/Wright

parameters

Read measurement concentration and
instrument status

Remark:

When connecting via RS-485
interface, a RS-232C↔RS-485
converter should be used.

A.4 Performance:

Repeatability: ±0.5% of full scale

Linearity: ±1% of full scale

Zero drift: ±1% of full scale/week

(±2% of full scale/week; range between 0
to 50ppm and 0 to 200ppm)
(±2% of full scale/day; below 0 to 50 ppm
range)

Span drift: ±2% of full scale/week

(±2% of full scale/day; below 0 to 50 ppm
range)