Up or down – Teledyne 7600 - Infrared gas analyzer - operating manual User Manual

Page 47

Advertising
background image

Infrared Gas Analyzer

Operation

Teledyne

Analytical

Instruments

47

4. In the “Calibration Parameters” screen that appears, point

the cursor to “Calibration Value” by pressing the UP or
DOWN key. Press the ENT key.


To change the concentration of a calibration gas:

1. Point the cursor to the channel you want to set by using

the

UP or DOWN

key. Press the ENT

key.

2. Select the concentration item you want to set by pressing

the UP or DOWN or SIDE key. Press the ENT key, and
the selected value is highlighted.

3. Enter the calibration gas concentration values (zero and

span). For any numerical value entry, press the UP or
DOWN key to increase or decrease a single digit. Select
the next digit by pressing the SIDE key.

4. When the concentration is correct, save the entry by

pressing the ENT key. The saved value becomes valid
from the next calibration process.